KM940 Hand-held Combustion Analyser

Strona 2 - CONTENTS

- 10 - Once the time has reached zero an audible beep will be heard and will show the selected fuel on the following display:- NATURAL GAS ∗P

Strona 3 - - 3

- 11 - 4.3 Main Displays The main display can be changed to show either 4 or 8 parameters at one time. Two options are available when 4 parame

Strona 4 - - 4

- 12 - 4.3.2 Line Scroll Mode Line scroll mode allows you to customise the display. Use the and keys to change the bo

Strona 5 - - 5

- 13 - 4.4 Sampling the Flue Gas Once the automatic calibration procedure has been completed and the specific fuel has been selected (See

Strona 6 - - 6

- 14 - 4.6 Regular Checks During Sampling Care must be taken at all times not to exceed the analysers operating specifications, in particular

Strona 7 - 1.4 Analyser Connections

- 15 - 4.8 Electromagnetic Compatibility The European Council Directive 89/336/EEC requires that electronic equipment does not generate elect

Strona 8 - 3. FIRST TIME USE

- 16 - 5. MOVING THROUGH THE MENUS 5.1 Basic Operation From the MAIN DISPLAY NETT C . . . . 0.0 O2 % . . . 20.9

Strona 9 - (44)-1707-375550

- 17 - 5.2 Menu Options and Settings 5.2.1 Main Menu The MAIN MENU consists of 4 sub menus which are shown below and detailed on the follow

Strona 10 - ∗PRESS -MENU- KEY∗

- 18 - Calculation of fuel constants are detailed in the Appendix. Fuel constants will have to be calculated before a user fuel can be enter

Strona 11 - - 11

- 19 - Selecting YES and will display the following screen. RESET SENSORS O2 % : 20.9 CO & NO = 0 PRESS ENTER MENU TO ESCAP

Strona 12 - - 12

- 2 - CONTENTS Page No: 1. ANALYSER LAYOUT AND FEATURES... 4-7 1.1 Instrument

Strona 13 - - 13

- 20 - 5.2.4 Display Menu LIGHT MODE CONTRAST : OFF : 8-PAGE : DEFAULT Allows the configuration of the display to be changed. LIGHT : Choo

Strona 14 - - 14

- 21 - Once an alarm has been exceeded the display will flash every two minutes warning the user of an alarm state and display the gas concent

Strona 15 - - 15

- 22 - Header : Allows two lines of 20 characters to be programmed into the analyser. The header appears on the top of the standard printout.

Strona 16

- 23 - 6. PRINTING INFORMATION Supplied as accessories for the KANE940 are an infra-red thermal printer or a dot matrix serial printer. Re

Strona 17 - - 17

- 24 - 7. STORING AND RETRIEVING DATA The KANE940 can store up to 100 combustion tests. Once stored, the data can be viewed on the display o

Strona 18 - - 18

- 25 - To print the data press . In the screen shown above locations 1 to 10 will be printed. During printing the following will b

Strona 19 - - 19

- 26 - 8. MAINTENANCE 8.1 Emptying and Cleaning the In-line Water Trap The in-line water trap should be checked and emptied on a regular bas

Strona 20 - - 20

- 27 - 9. PROBLEM SOLVING The following is a list of problems that may occur on the instrument through its operating life. If the cause of the

Strona 21 - - 21

11. PRODUCT SPECIFICATION Parameter Resolution Accuracy Range Temp Measurement Flue Temperature with probe 1.0ºC/F +2.0ºC +0.3% reading 0-600º

Strona 22 - - 22

- 29 - Dimensions Weight Handset Probe 1kg 220mm x 55mm x120mm L 420mm x Dia 8mm with stainless steel shaft, type K thermocouple and 3m hos

Strona 23 - ∗∗∗∗∗ Printing ∗∗∗∗∗

- 3 - 8. MAINTENANCE ...26 8.1 Emptyi

Strona 24 - - 24

- 30 - APPENDICES A - Main Display Parameters The parameters and their meanings are detailed as follows : - DATE : Analyser date. See Set-U

Strona 25 - - 25

- 31 - (Ta) temperature calculation if an INLET probe is not fitted. CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by cal

Strona 26 - 8. MAINTENANCE

- 32 - B. COMBUSTION EFFICIENCY CALCULATION The efficiency calculation is based upon British Standard BS845. This identifies three sources o

Strona 27 - 10. ANNUAL RE-CALIBRATION

- 33 - Calculated data: Tnet = Net Temperature % CO2 content in flue gas % Dry Flue Gas losses % Wet losses % Unburned carbon l

Strona 28 - 11. PRODUCT SPECIFICATION

- 34 - C. CALCULATION OF FUEL DATA For any fuel not specified by Kane International the net calorific value, gross calorific value and compo

Strona 29 - - 29

- 35 - D. ELECTROMAGNETIC COMPATABILITY (CE) STATEMENT This product has been tested for compliance with the following generic standards

Strona 30 - APPENDICES

- 4 - 1. ANALYSER LAYOUT AND FEATURES 1.1 Instrument Features and Keypad ON/OFF UP Scrolls up through

Strona 31 - - 31

- 5 - 1.2 Instrument Layout (Rear) SENSORS FITTEDCharger SocketNOTE! - CoverDoes Not RemoveSerial NumberOptions LabelExhaust Port*NOTE! Do not

Strona 32 - - 32

- 6 - 1.3 Standard Probe Configuration

Strona 33

- 7 - 1.4 Analyser Connections Gas Connectionto InstrumentParticle FilterIn-line Water TrapNeoprene HoseFlue TemperatureConnector

Strona 34 - - 34

- 8 - 2. SAFETY WARNING This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted

Strona 35 - - 35

- 9 - 4. NORMAL START UP SEQUENCE 4.1 Every Time You Use The Analyser BEFORE SWITCH-ON CHECK THAT: the particle filter is not dirty t